QUICK ANSWER
IMU data sheets mix units: °/h and °/s, °/√h and (°/s)/√Hz, µg and m/s². The fixed factors are simple: 1 °/√h equals 1/60 (°/s)/√Hz, and 1 (m/s)/√h equals about 1,700 µg/√Hz. The trap is the noise density. Measured as a one-sided spectrum it is √2 times larger than the Allan coefficient behind the ARW. The converter handles both conventions.
01
Which conversion factors do I need most often?
| From | To | Factor |
|---|---|---|
| 1 °/h | µrad/s | 4.848 |
| 1 °/h | °/s | 2.778 × 10⁻⁴ |
| 1 °/√h | (°/s)/√Hz, Allan coefficient | 1/60 = 0.01667 |
| 1 (°/h)/√Hz, Allan coefficient | °/√h | 1/60 = 0.01667 |
| 1 µg | m/s² | 9.80665 × 10⁻⁶ |
| 1 µg | tilt | ≈ 1 µrad = 0.206″ |
| 1 (m/s)/√h | µg/√Hz, Allan coefficient | 1,699.5 |
| 1 mGal | µg | 1.0197 |
Standard gravity g = 9.80665 m/s².
02
Why is the noise density not simply ARW × 60?
The angle random walk is defined through the Allan variance: σ²(τ) = N²/τ. In this definition N² is a two-sided power spectral density. A spectrum analyzer shows a one-sided density, which carries the same power in positive frequencies only and is therefore twice as large. Its square root, the noise density on many data sheets, is √2 times N.
The Epson data sheets confirm the order of magnitude: across four models the stated noise density is 1.2 to 1.5 times ARW × 60 (own calculation, see How to read an IMU datasheet). If a manufacturer derives the density directly from the Allan deviation at τ = 1 s, the factor is exactly 60. Check which convention applies before you compare two sensors.
03
How do I get the RMS noise for my bandwidth?
σRMS = one-sided noise density × √(equivalent noise bandwidth)
For an ideal filter the equivalent noise bandwidth equals the cut-off frequency. For a first-order low-pass it is 1.571 times, for a second-order Butterworth 1.111 times the −3 dB frequency. The lower table in the tool calculates this for the values you entered, together with the angle, velocity and position errors after a time of your choice.
To see how these errors add up in navigation, use the IMU drift simulator. To derive the coefficients from your own measurement, use the Allan deviation calculator.
QHow do I convert °/√h to (°/s)/√Hz?
Divide by 60. An ARW of 0.06 °/√h is 0.001 (°/s)/√Hz as Allan coefficient, or about 0.0014 (°/s)/√Hz as one-sided noise density.
QHow do I convert µg/√Hz to (m/s)/√h?
Multiply by 9.80665 × 10⁻⁶ × 60, that is about 5.88 × 10⁻⁴. If the value is a one-sided noise density, divide by √2 first.
QWhat does 1 µg of accelerometer bias mean for tilt?
About 1 µrad, or 0.2 arc seconds. A bias of 24 µg corresponds to a tilt error of 0.0014°.
QIs °/√h the same as (°/h)/√Hz?
No. 1 °/√h equals 60 (°/h)/√Hz. The first is a random-walk coefficient per square root of an hour, the second a rate density per square root of bandwidth.